MicroDock II
User Manual
The cursor automatically displays below the first value of the
month. Each value is selected, changed, and bypassed individu-
ally.
C
2. Press
to toggle between 0 or 1. When the correct value
displays, press
C
sel to confirm the change. The cursor then
automatically moves to the second value of the month.
Or
Press
C
sel to bypass the first value. The base station automat-
ically retains the current value and proceeds to the second value
of the month.
3. Repeat step #2 for the remaining time and date values.
4. Press
C
exit to return to the user options menu.
C
5. Press
again to return to normal operation.
exit
Inlet Setup
Inlet Setup defines the
•
the gas type(s)
•
the corresponding gas concentrations level(s), and
•
the gas cylinder lot # of each selected gas.
40
Failed tests can result if the inlets are not setup correctly.
For initial base station activation, ensure the inlets are installed correctly
as follows:
•
Inlet 1 is the default connection for ambient air.
•
Inlet 2 is the default connection for four-gas mix, unless other-
wise specified at purchase.
•
Inlets 3-5 are designed connections for additional gases. How-
ever, unless specified at the time of purchase, inlets 3-5 are con-
figured for ambient air.
To confirm that the inlets are setup correctly, complete the following:
1. From the user options menu, press
the Inlet Setup option.
2. Press
C
sel to confirm the selection and access the inlet 1
(ambient air) screen.
Inlet 1 is designated for ambient air only and cannot be
changed.
3. From the inlet 1 screen, press
a Warning
C
or
C
Note
C
to access the inlet 2 screen.
to scroll to