Inserting the GasAlertClip Extreme and the GasAlert
Extreme
Infrared or intense ambient light (sun or halogen) may interfere
with the base station/detector communication.
Figure 5. Inserting the GasAlertClip Extreme and the
GasAlert Extreme
12
a Caution
Table 7. Inserting the GasAlertClip Extreme and the GasAlert
Item
1
Alligator clip
2
Docking module lid
3
Release tabs
4
Detector bay
If the GasAlertClip Extreme or GasAlert Extreme is fitted with a
calibration cap, it must be removed prior to inserting the detec-
tor into the docking module.
1. Activate the detector and wait until it is in normal operation.
2. Ensure the alligator clip is closed and the ring is resting flat to
prevent disruptions with the data transmission.
3. Press the two release tabs on the docking module and raise the
lid.
4. Lower the detector (serial number label up) into the detector bay.
Push forward to ensure the top of the detector connects with the
top of the bay.