CHARACTERISTICS
CHARACTERISTICS

2.1.12 Signal delay

VISUAL SIGNAL DELAY
DELAY BETWEEN
CHANNELS
CH1 and CH2
CH3 and CH4
CH1...CH4
Any two channels

2.1.13 Vertical accuracies

ACCURACY
deflection factor
Gain error (d.c.)
Non linearity
MAX. BASELINE
INSTABILITY
Jump (all between steps,
VAR and N/l)
Drift
Temperature
coefficient
CHANNEL ISOLATION
Of deselected channels
at 10 MHz
Of deselected channels
at upper transition point
Between selected channel
CMRR
at 1 MHz
at 50 MHz
Note 1:
add 1.5 % for variable gain mode
Note 2:
2 div centerscreen signal with a frequency of 50 kHz, shifted within central 6 div.
Note 3:
At 10 MHz; input to deselected channel equivalent to 8 div. or less.
Note 4:
PM3092/3094 a t200 MHz; PM3082/3084 at 100 MHz; channels with equal V/div setting;
input to deselected channels equivalent to 6 div.
Note 5:
PM3092/3094 at 200 MHz; PM3082/3084 at 100 MHz; channels with equal V/div.
settings; input to either channel 6 div. PM3092/94: max input amplitude 3 Vpp
Note 6:
Between any two input channels at same attenuator setting; VAR of V/div setting adjusted
for best CMRR; measured with max. 8 div input at each channel (center screen).
SPECIFICATION
>15 ns
>13 ns
<250 ps
<250 ps
<250 ps
<500 ps
±1.3 %
<2 %
0.2 div or 1 mV
0.1 div/h
0.03div/K
100:1
50 :1
50 :1
100:1
25:1
ADDITIONAL INFORMATION
PM3092, PM3094
PM3082, PM3084
4 channel instruments
2+2 channel instruments
Over central 6 div.
See note 1
See note 2
Whichever is greater
(after autocal)
See note 3
See note 4
See note 5
See note 6
2 - 5