6.2

RECOMMENDED TEST EQUIPMENT

Note:
The digital multimeter and oscilloscope are not required for this test.
Note:
The FLUKE 5500A Multiproduct Calibrator is always used in SCOPE mode. The output
signal must be taken from the BNC output 'Scope'.
Type of instrument
Function generator
Constant amplitude
sine wave generator
Square-wave
calibration generator
Time marker generator
Digital multimeter
Variable voltage
transformer (VARIAC)
TV pattern generator with
video output
Oscilloscope
BNC/Probe tip adapter
500 cables,
750 cable,
500 terminations,
750 termination,
10:1 attenuator,
T-piece,
power splitter
2:1 attenuator
Required specification
Freq: 10 Hz... 100kHz
Sine wave/square-wave
Ampl: 0...20 V(pp)
Rise time <30 ns
Duty cycle 50%
Freq: 50 kHz...300 MHz.
Constant pp. amplitude
of 10 mV to 5.5V.
For ampl. calibration:
Freq: 1 kHz
Ampl: 10 mV...50 V
For rise time
measurements:
Freq: 10 kHz... 1 MHz
Ampl: 10...900 mV
Rise time: <1 ns
Repetition rate:
0.5s...2 ns
Wide voltage and
current ranges.
Well insulated output
voltage 90...264V (ac)
The bandwidth must be
the same or higher
than the bandwidth of
the instrument under test.
For bandwidth check
in PM3092
Fluke or e.g. Tektronix
BNC types for fast risetime
square-wave and
high-frequency sinewave
applications.
PERFORMANCE TEST
Example of recommended
instrument
Fluke 5500A mode: wavegen
Fluke 5500A mode: levsine
Fluke 5500A mode: volt
Fluke 5500A mode: edge
Fluke 5500A mode: marker
Fluke 29/79 with AC, DC
and resistance ranges.
High voltage probe.
Required: 1% accuracy,
model 80K-40
Fluke PM5418
Fluke PM3094
ord.nr. 5322 263 50022
Fluke PM9074
Fluke PM9075
Fluke PM9581
TEK 011-0055-01
TEK 011-0059-02
Fluke PM9067
Fluke PM9584/02
TEK 011-0069-02