8 - 6
8.5.6
Removal of Final XYZ amplifier unit A2
ATTENTION: On the XYZ unit there are parts that carry high voltages. If working on the unit under
live condition cannot be avoided, it must be done by a qualified technician who is
aware of the dangers involved.
-
Remove the screw that secures (and grounds) the unit to the bottom chassis plate.
-
Disconnect the 4 wires that lead to the CRT (The X- and Y-deflection plates): this action must be
done carefully to avoid damage to the side connections of the CRT. For correct reinstallation refer
to the wiring diagram in chapter 4.2.
-
Bend out the two clamping lips that secure the unit at the top side. The unit is loosened now.
NOTE: The unit can be put now in an inclined position as shown in figure 8.3. Measuring on the SMD-
component side in working condition is possible then. Measuring the output wires that lead
to the X- and Y-deflection plates is possible with a 10 fcQ damping resistor between
measuring point and probe tip. This avoids oscillations.
-
Unplug two ribbon cables and take the unit out of the chassis.
-
Unplug the delayline connector.
Figure 8.16 Final XYZ amplifier in inclined position
CORRECTIVE MAINTENANCE PROCEDURES