6.3.40 Probe Adjust signal; frequency and output voltage

The Probe Adjust (CAL) signal is a calibration signal with fixed frequency and voltage. In this test, the
values of frequency and voltage are checked.
test equipment:
None
test set-up:
OSCILLOSCOPE UNDER TEST
CABLE
settings/procedure:
1 -
Connect the Probe Adjust signal to input CH1 and press the AUTOSET key.
2 -
Select GND of CH1.
3 -
Set the trace in the center of the screen.
4 -
Switches GND of CH1 off.
5 -
Select DC input coupling for CH1.
requirements:
1 -
Verify that a positive going square-wave signal of 0.6 V(pp) is displayed, i.e. 3 divisions vertical
at 0.2V.
2 -
Verify that the frequency of the displayed signal is about 2 kHz, i.e. a period time between
4.0 ... 6.0 divisions horizontal at MTB 0.1 ms.