Specifications; Educational Services; Installation - Emerson Fisher 4200 Instruction Manual

Electronic position
Table of Contents

Advertisement

Instruction Manual
D200354X012
Table 4. Fisher 4211 and 4221 Position Transmitter EMC Summary Results—Immunity
PORT
Electrostatic Discharge (ESD)
Radiated EM field
Enclosure
Rated power frequency magnetic
Burst (fast transients)
I/O signal/control
Specification limit = ±1% of span
1. A = No degradation during testing. B = Temporary degradation during testing, but is self‐recovering.
For instruments equipped with position switches, the high relay position switch can be set at any point of travel. The
low relay position switch can be offset from the high relay position switch trip point by as little as 1/20 of the travel. In
the event of a power loss both position switch relays de‐energize, indicating loss of transmitter power.

Specifications

Specifications for the 4200 transmitters are listed in table 1.
WARNING
This product is intended for a specific current range, temperature range and other application specifications. Applying
different current, temperature and other service conditions could result in malfunction of the product, property damage or
personal injury.

Educational Services

For information on available courses for 4200 electronic position transmitters, as well as a variety of other products,
contact:
Emerson Process Management
Educational Services, Registration
P.O. Box 190; 301 S. 1st Ave.
Marshalltown, IA 50158-2823
Phone: 800-338-8158 or
Phone: 641-754-3771
FAX: 641-754-3431
e‐mail: education@emerson.com

Installation

WARNING
Always wear protective clothing and eyewear when performing any installation operations to avoid personal injury or
property damage.
PHENOMENON
BASIC STANDARD
field
Surge
Conducted RF
4 kV contact
IEC 61000‐4‐2
8 kV air
80 to 1000 MHz @ 10V/m with 1 kHz AM at 80%
IEC 61000‐4‐3
1400 to 2000 MHz @ 3V/m with 1 kHz AM at 80%
2000 to 2700 MHz @ 1V/m with 1 kHz AM at 80%
IEC 61000‐4‐8
60 A/m at 50 Hz
IEC 61000‐4‐4
1 kV
IEC 61000‐4‐5
1 kV (line to ground only, each)
150 kHz to 80 MHz at 3 Vrms
IEC 61000‐4‐6
1 kHz AM at 80%
4200 Transmitters
TEST LEVEL
March 2011
PERFORMANCE
(1)
CRITERIA
A
A
A
A
B
A
5