Wash carts, modules and inserts
Wash carts, modules and inserts should be checked daily to make
sure they are functioning correctly. The washer-disinfector is supplied
with a check list.
Check the following points:
– Are the rollers in proper condition, and are they securely attached
to the wash cart/basket?
– For wash carts in the modular system, are the caps in the module
connection working properly?
– Are all spray jets, spray sleeves and hose adaptors securely
attached to the wash cart/insert?
– Are all spray jets, spray sleeves, and hose adapters unclogged so
that wash water can flow through?
– Are all caps, covers, and fasteners securely attached to the spray
sleeves?
Where applicable:
– Do the spray arms rotate freely?
– Are the spray jets clogged? See the section on "Maintenance/
Cleaning the spray arms".
– Are the screws in the magnetic rails for automatic program
recognition tightly secured?
– Is the magnetic strip of the automatic program recognition feature
free of attached metal objects?
Wash carts, baskets and inserts must be inspected in the course of
the routine checks of the washer-disinfector after 2000 operating
hours or at least once per year, see "Maintenance/Routine checks".
Maintenance