Program Sequence; End Of Program; Cancelling A Program - Miele G 7825 Operating Instructions Manual

Washer-disinfectors
Table of Contents

Advertisement

Operation
In operating level
B or D
46

Program sequence

The program will start automatically as soon as the Start button is
pressed.
The program stages are shown in the display during the program
sequence.
Detailed information on program sequence is given in the appendix
of the Programming manual.
 Do not change the printer paper roll or ribbon cartridge during a
running program.

End of program

At the end of the program PROGRAMME - END is shown in the
display and the background lighting flashes.
Press any button to stop the lighting from flashing.
For general information on how to switch off the flashing signal, see
the section "System functions" in the Programming Manual.

Cancelling a program

A program can only be interrupted or cancelled in operating levels
B and D.
In order to conduct a performance qualification to assess the
cleaning result, the program must be cancelled before it proceeds
to the disinfection step (as per EN ISO 15883-1; CAN/CSA
Z.15883-1-09).
Press the Stop button .
The program is interrupted.
The following will appear in the display:
CANCEL OR>CONTINUE<
 In cases involving mandatory decontamination, disinfectant
must be added to the contaminated water before it can be
discharged into the sewerage system.
The door on the unclean side can be opened for this purpose.
Use cursor  to select >CANCEL<, the cursors will start flashing.
Press the Start button .
The program is cancelled and the water drained away.
WATER DRAIN appears in the display.