Application technology
26

OR shoes

OR shoes and/or insoles made of heat-sensitive materials should be
thermo-chemically cleaned and disinfected at 60°C using the CHEM-
DESIN program.
A thermal disinfection program (SHOE-TD-75/2 program) can be used
as long as the manufacturer has confirmed the thermo-stability of the
product(s) involved.
For reprocessing OR shoes to an A
SHOE-60 program must be programmed in retrospectively by Miele.
For information on the efficacy of thermochemical disinfection,
contact the disinfectant manufacturer directly.
OR shoes should be cleaned and disinfected in a machine
installed specifically for this purpose only. If OR shoes are to be
reprocessed in a washer-disinfector which is used for other
applications, a risk assessment must be carried out by the user.
The wash cart E 775-2 can be used with a suitable insert, e.g. E 930.
When OR shoes are cleaned, large quantities of lint can
accumulate. For this reason, be sure to check the wash cabinet
filters frequently and clean as needed (see section on "Cleaning
and care").
=60 disinfection standard, the