Overshoot .............................................................. <10 %
....................................... 140 dB, dc to 400 Hz
Common Mode Rejection
Frequency
Ranges (Hz) ........................................................ 10.000 - 119.99
Absolute Specification ........................................ ±0.0025 %
Resolution ........................................................... 11.999 counts
Phase Lock (Selectable Rear Panel BNC Input)
Phase Specification (except 1100 V range) ........ >30 Hz: ±1 ° + 0.05 °/kHz), <30 Hz: ±3 °
Input Voltage ....................................................... 1 V to 10 V rms sine wave (do not exceed 1 V for mV ranges)
Frequency Range ............................................... 10 Hz to 1.1999 MHz
Lock Range ......................................................... ±2 % of frequency
Lock-In Time ....................................................... Larger of 10/frequency or 10 msec
Phase Reference (Selectable Rear Panel BNC Output)
Range ................................................................. ±180 °
Phase Absolute Specification ............................. (except 1100 V range) ±1 ° at quadrature points (0 °, ±90 °, ±180 °)
Stability ............................................................... ±0.1 °
Resolution ........................................................... 1 °
Output Level ....................................................... 2.5 V rms ±0.2 V
Frequency Range ............................................... 50 Hz to 1 kHz, usable 10 Hz to 1.1999 MHz

Resistance Specifications

Absolute Specification of Characterized Value
Nominal Value
(Ω)
24 Hours
0
50 μΩ
1
85
1.9
85
10
23
19
23
100
10
190
10
1 k
7
1.9 k
7
10 k
6
19 k
6
100 k
7
190 k
8
1 M
13
1.9 M
15
10 M
33
19 M
43
100 M
100
0.1200 k - 1.1999 k
1.200 k - 11.999 k
12.00 k - 119.99 k
120.0 k - 1.1999 M
elsewhere ±2 °
5730A Resistance Specifications
±5 °C from calibration temperature
90 Days
180 Days
99 % Confidence Level
50 μΩ
50 μΩ
95
100
95
100
25
26
25
26
11
11.5
11
11.5
7.2
7.5
7.2
7.5
7
7.5
7
7.5
8
9
10
11
14
14.5
17
19
37
40
47
50
110
115
[1]
1 Year
24 Hours
±ppm
50 μΩ
50 μΩ
110
32
110
25
27
5
27
4
12
2
12
2
8
2
8
2
8
2
8
2
10
2
12
2
15
2.5
21
3
46
10
55
20
120
50
Multifunction Calibrator
Electrical Specifications
Relative ±1 °C
90 Days
50 μΩ
40
33
8
7
4
4
3
3
3
3
3
3
5
6
14
24
60
31

Hide quick links: