Electrical Specifications - Fluke 5730A Getting Started Manual

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Getting Started Manual

Electrical Specifications

The product specifications describe the Absolute Instrumental Uncertainty of the Product. The product specifications
include stability, temperature, and humidity; within specified limits, linearity, line and load regulation, and the reference
standard measurement uncertainty. The product specifications are provided at a 99 %, k=2.58, normally distributed and a
95 %, k=2, normally distributed level of confidence. Fluke Calibration guarantees product performance to the 99 % level of
confidence.
The relative specifications are provided for enhanced accuracy applications. The specifications apply when range
constants are adjusted (see "Range Calibration"). To Calculate an enhanced absolute specification from the relative
accuracy specification, it is necessary to combine the uncertainty of your external standards with the pertinent relative
specifications.
Specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the Product has been turned off.
DC Voltage Specifications
Range
99 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
95 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
Notes:
Perform the DC Zero calibration every 30 days. In addition, perform the DC Zero calibration after powering up the unit the first time after
unpacking following a shipment or if exposed to an environmental change of greater than 5 °C.
1.
For radiated EMI fields >400 MHz and <500 MHz, add 1 ppm.
DC Voltage Secondary Performance Specifications and Operating Characteristics
Range
220 mV
2.2 V
11 V
22 V
220 V
1100 V
Notes:
1.
Stability specifications are included in the absolute specification values in the primary specification tables.
Temperature coefficient is an adder to accuracy specifications that does not apply unless operating more than ±5 °C from
2.
calibration temperature.
24
5730A DC Voltage Specifications
Absolute / ±5 °C from calibration temperature
Resolution
24 Hours
10 nV
5 + 0.5
100 nV
3.5 + 0.8
2.5 + 3
1 μV
1 μV
2.5 + 5
10 μV
3.5 + 50
5 + 500
100 μV
10 nV
4 + 0.4
100 nV
3 + 0.7
2 + 2.5
1 μV
1 μV
2 + 4
10 μV
3 + 40
4 + 400
100 μV
[1]
±1 °C
Stability
24 Hours
10 - 40 °C
±(ppm output + μV)
0.3 + 0.3
0.4 + 0.1
0.3 + 1
0.3 + 0.1
0.3 + 2.5
0.15 + 0.2
0.4 + 5
0.2 + 0.4
0.5 + 40
0.5 + 200
90 Days
180 Days
±(ppm output
7 + 0.5
8 + 0.5
4 + 0.8
4.5 + 0.8
3 + 3
3.5 + 3
3 + 5
3.5 + 5
4 + 50
5 + 50
6 + 500
7 + 500
6 + 0.4
6.5 + 0.4
3.5 + 0.7
4 + 0.7
2.5 + 2.5
3 + 2.5
2.5 + 4
3 + 4
3.5 + 40
4 + 40
4.5 + 400
6 + 400
Temperature Coefficient
[2]
Adder
0 - 10 °C and
40 - 50 °C
±(ppm output + μV) / °C
1.5 + 0.5
1.5 + 2
1 + 1.5
1.5 + 3
0.3 + 5
1.5 + 40
0.5 + 10
3 + 200
Relative ±1 °C
1 Year
24 Hours
[1]
+ μV)
9 + 0.5
2 + 0.4
6 + 0.8
2 + 0.8
4 + 3
1 + 3
4 + 5
1 + 5
6 + 50
2 + 50
8 + 500
2.5 + 400
7.5 + 0.4
1.6 + 0.4
5 + 0.7
1.6 + 0.7
3.5 + 2.5
0.8 + 2.5
3.5 + 4
0.8 + 4
5 + 40
1.6 + 40
6.5 + 400
2 + 400
Noise
Linearity ±1 °C
Bandwidth
0.1 - 10 Hz
pk-pk
±(ppm output + μV)
1 + 0.2
0.15 + 0.1
1 + 0.6
0.15 + 0.4
0.3 + 2
0.15 + 2
0.3 + 4
0.15 + 4
1 + 40
0.15 + 60
1 + 200
0.15 + 300
90 Days
2.5 + 0.4
2.5 + 0.8
1.5 + 3
1.5 + 5
2.5 + 50
3 + 400
2 + 0.4
2 + 0.7
1.2 + 2.5
1.2 + 4
2 + 40
2.4 + 400
Bandwidth
10 - 10 kHz
RMS
μV
5
15
50
50
150
500

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents