A40B
Instruction Manual

Resistance Specifications

Shunt Nominal
Current
1 mA
10 mA
20 mA
50 mA
100 mA
200 mA
500 mA
1 A
2 A
5 A
10 A
20 A
50 A
100 A
[1]
Stability specification combines long term change due to aging (permanent) and short term fluctuation due to humidity
changes when shunts are used and stored within specified humidity limits. Add 20 (±μΩ/Ω) if humidity is >50 % RH).
[2]
Assume rectangular distribution when combining with other uncertainty contributions.
[3]
Calibrated resistance values include the effects of power coefficient at the nominal current. For currents other than nominal,
apply an additional uncertainty for the power coefficient from:
Adder
6
Maximum
Deviation
Nominal
from Nominal
Resistance
Resistance
(Ohms)
[2]
(±μΩ/Ω)
800
250
80
250
40
250
16
250
8
250
4
250
1.6
250
0.8
250
0.4
250
0.16
250
0.08
250
0.04
250
0.016
250
0.008
250
=
Power
_
Coefficien
t
_
Multiplier
Approximate
Uncertainty of
12 Month
Calibrated
Stability
Value at 95 %
[1][2]
Confidence
(±μΩ/Ω)
(±μΩ/Ω) TCal
±1°C
8.2
18
6.8
18
8.2
18
8.3
18
8.3
18
8.6
18
9.6
18
9.3
18
9.4
18
9.9
18
15
18
14
18
24
18
28
18
2
I
 
 
Applied
×
1
I
No
min
al
Power
Temperature
Coefficient
Coefficient
Multiplier
[2]
(±μΩ/Ω/ºC)
[2][3]
(±μΩ/Ω)
5
1
2.5
1
4.5
1
4.5
1
2.5
2
3.5
4
4.5
13
4.5
26
4.5
26
4.5
30
4.5
65
4.5
78
4.5
105
4.5
105