Bottle Mode
Page 96
11. When the backflush and equalization is completed, the test summary dialog box
appears, with the results of the calibration and a notification of whether the calibration
failed or passed. The results also appear on the printout if a printer is attached. See
Figure
77.
Summary text lists the results of the test. The results appear as they would
appear on the printed report.
Export icon takes you to the export dialog box, where you can save the results of
the test to a USB flash drive.
Checkmark icon accepts the calibration, if the values passed, and takes you to
the unload/replace bottles dialog box.
Cancel icon rejects the calibration results and takes you to a dialog box that asks
if you want to rerun the test. If you touch the forward icon, the analyzer returns you
to the beginning of the test. All previous data are discarded.
Auto accept. If you have selected to auto accept the results and return online, the
results are accepted and sent to the printer, and the analyzer returns to the home
screen. If you have selected manually accept the results and return online, the
analyzer returns to the home screen immediately following the completion of the
test. If you have selected manually accept the results and remain offline, the test
summary dialog box appears.
12. When the unload/replace bottles dialog box appears, follow the instructions for
unloading or replacing bottles, then press the done icon. See
Figure 77 Test summary dialog box
Figure
78.