Bottle Mode
Page 94
C of A value text box allows you to enter the certified analysis value for the
standard that is being used.
Volume text box allows you to enter the volume of water, in mL, for the specified
standards bottle.
Expiration date text box allows you to enter an appropriate expiration date, in
ISO yyyy-mm-dd format, and time, in hh:mm:ss format, for the specified TOC
calibration standard.
Forward icon takes you to the review test setup dialog box. See
Cancel icon returns you to the load bottles dialog box.
Date text box allows you to enter the date the test will run. The default value is
tomorrow's date. You may enter any valid date, in ISO "yyyy-mm-dd" format, that
is within three days from the loading of the standards.
Time text box allows you to enter the time the test will run. The default time is the
current time. You may enter any valid time in hh:mm:ss format.
Back icon takes you to the previous dialog box. All changes made in this dialog
box are retained.
Finish icon takes you back to the home screen. All changes made in any of the
dialog boxes are retained and the test is scheduled to run. Prior to exiting this
dialog box, the analyzer verifies that the scheduled date and time is within three
days of the current date and time. It also verifies that none of the bottles will expire
prior to the scheduled date.
10. When the test running dialog box appears, the analyzer begins running the test. See
Figure
75.
Figure 74 Schedule test dialog box
Figure
72.