Bottle Mode
Page 92
5. To use default settings, touch the use default settings radio button in the TOC
calibration setup dialog box. To run any other configuration, touch the use custom
settings radio button, then define the desired configuration. (See
The following procedure relies on the recommended default setting.
6. Touch the forward icon. The OASIS door will unlock for five seconds. If the door is
not opened within five seconds, the unlock icon can be used to unlock the door
manually.
7. Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottles in the correct positions.
See
Figure
71. From left:
Bottle 1 = blank
Bottle 2 = 250 ppb sucrose
Bottle 3 = 500 ppb sucrose
Bottle 4 = 750 ppb sucrose
Figure 71 Load bottles dialog box, TOC calibration
Bottle positions graphics show you the order in which to install the bottles.
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon takes you to the run standards dialog box.
Forward icon takes you to the review test setup dialog box if RFID tagged bottles
are being used (see
bottles that are not RFID tagged are being used (see
Cancel icon returns you to the previous dialog box. All changes made in any of
the dialog boxes are lost.
8. Close the bottle bay door.
9. Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
have RFID tags, the enter bottle information dialog box appears. See
Figure
72), or to the enter bottle information dialog box if
Figure 70, page
Figure
73).
Figure
72. If the bottles do not
Figure
73.
91.)