Figure 84 Load bottles dialog box, TOC validation
Bottle positions graphics show you the order in which to install the bottles.
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon takes you to the previous dialog box.
Forward icon takes you to the review test options dialog box.
Cancel icon returns you to the run standards dialog box. All changes made in any
of the dialog boxes are lost.
8. Close the bottle bay door.
9. Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
Otherwise, the enter bottle information dialog box appears. See
10. When the test running dialog box appears, the analyzer begins running the test.
After verifying the presence of the appropriate standards, the analyzer performs a
temperature stabilization. Once the temperature has stabilized, the blank and
standards bottles are analyzed in the proper sequence.
Once all the bottles have been analyzed, the analyzer compares the current
results to the standard calibration. If the average of each standard is within ±15%
of the expected value, the validation is accepted. If any of the averaged values
are greater than ±15% of the expected value, the validation is rejected.
The dialog box displays the current status of the test and lists the results of each
repetition and bottle as it is running.
11. When the validation is completed, the test summary dialog box appears, with the
results of the validation and a notification of whether the validation failed or passed.
The results also appear on the printout if a printer is attached. See
page
96. Touching the forward icon takes you to the run again dialog box
page
97). If the test failed or you have rejected the calibration results, at the run
again dialog box, touch the checkmark icon to run the validation again or the cancel
icon to cancel the run.
Note: Immediately upon completion of the test the bottles are backflushed and allowed to equilibrate
to ambient pressure. DO NOT ATTEMPT TO REMOVE THE BOTTLES UNTIL THE BACKFLUSH
AND PRESSURE EQUALIZATION IS COMPLETE.