Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Run test icon runs the plumbing test.
Cancel icon returns you to the previous dialog box. All changes made in any of
the dialog boxes are lost.
3. After you've entered the correct bottle information, touch the run test icon, which
returns you to the plumbing test dialog box. See
4. Select the desired test from the flow test selection drop down list box.
5. When the test is completed, touch the unload bottles icon. The unload/replace bottles
dialog box appears. See
Figure 111 Unload/replace bottles dialog box
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Done icon returns you to the plumbing test dialog box.
6. If a test is chosen immediately, the analyzer asks if you wish to install a bottle. If yes
is selected, the load bottles dialog box illustrates the appropriate port in which to
install the bottle. The enter bottle information screen requires you to enter the bottle's
volume. The next button then starts the test.
Figure 108, page
Figure
111.
Diagnostics
131.
Page 133