Figure 80 Load bottles dialog box, conductivity calibration
Bottle positions graphics show you the order in which to install the bottles.
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon takes you to the run standards dialog box.
Forward icon takes you to the review test options dialog box.
Cancel icon returns you to the previous dialog box. All changes made in any of
the dialog boxes are lost.
12. Close the bottle bay door.
13. Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
Otherwise, the enter bottle information dialog box appears. See
14. If the selection has been made to use the conductivity meter verification resistor,
install the resistor onto the analyzer's 9-pin printer/calibration port, as illustrated in
Figure
81.
If the calibration is being performed on a schedule, the conductivity meter test
dialog box does not appear, and the calibration occurs automatically.
Otherwise, the conductivity meter test dialog box appears when the analyzer
passes the conductivity meter test. See
Figure 72, page
Figure 73, page
Figure
82.
Bottle Mode
93.
93.