Advertisement

To avoid damaging the analyzer:
Do not leave water inside the analyzer before putting it into storage.
Do not expose the analyzer to temperature extremes.
To purge the PAT700 of entrained water:
1. Remove the analyzer from service and disconnect it from the water system.
2. With the analyzer inlet and outlet water connections open to atmosphere, apply
power to the analyzer.
3. In the TOC setup dialog box under the general tab, select "use pump".
4. At the home screen, touch the << icon to open the sliding toolbar.
5. Touch the diagnostics icon.
6. Touch the tests tab, then touch the plumbing test icon.
7. Follow the procedure described in
the analyzer (flow through cell with pump and sample from all four bottles locations).
8. If necessary, repeat
Step 4
analyzer.
Maintenance and Troubleshooting
Plumbing test, page
through
Step 7
until all water has drained from the
131, to drain the water from
Page 175