Hach Anatel PAT700 User Manual page 156

Total organic carbon analyzer
Hide thumbs Also See for Anatel PAT700:
Table of Contents

Advertisement

Modbus Protocol
Page 154
Configuration Block
The Configuration Data block comprises parameters that directly affect the sampling
characteristics of the instrument. Any modifications to these registers will restart the
current sample if a sample is active.
Note: Use FC16 where '*' is indicated.
Address
Register Description
0100
Instrument IP address
0102
Subnet Mask
0104
Gateway
0106
DNS Server
0108
WINS Server
0110
DHCP Flag
0111
Language identification code
0112
Mode of Operation
0113
Current State
0114
Cycle Time
0116
Flush Time
0118
Idle Mode
0119
Bypass flow
0120
Clean time mode
0121
Clean time
0123
UTC Date and Time
0127
UTC Time format
0128
Pump sampling
0129
Digital Control
0130
Temperature display unit
0131
Conductivity display unit
0132
Cell calibration gain
0136
Cell calibration offset
0140
User TOC Slope
0144
User Conductivity Slope
0148
Thermistor coefficient A
0152
Thermistor coefficient B
0156
Thermistor coefficient C
0160
TOC Print
0161
TOC Print Percent change
0162
Conductivity Print
0163
Conductivity Print Time
0165
Conductivity Print Percent change
0166
TOC alarm
0167
TOC alarm limit
0168
Conductivity alarm (uncompensated only)
0169
Last alarm
Configuration Block
Size
Access
(Words)
R
2
R
2
R
2
R
2
R
2
R
1
R/W
1
R/W
1
R
1
R/W*
2
R/W*
2
R/W
1
R/W
1
R/W
1
R/W*
2
R/W*
4
R/W
1
R/W
1
R/W
1
R/W
1
R/W
1
R
4
R
4
R
4
R
4
R
4
R
4
R
4
R/W
1
R/W
1
R/W
1
R/W*
2
R/W
1
R/W
1
R/W
1
R/W
1
R
1