Internal Triggering; Software (Bus ) Triggering - Agilent Technologies 34410A User Manual

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

• The configured range mode (auto or manual) will be used for reading
• To enable reading hold, press

Internal Triggering

The internal triggering mode is available from the remote interface only.
In the internal trigger mode, the trigger signal is always present. When
you place the multimeter in the wait–for–trigger state, the trigger is issued
immediately. This is the default trigger source for remote interface
operation.
• Remote Interface Operation: The following command selects the internal

Software (Bus ) Triggering

The bus trigger mode is available from the remote interface only.
The bus trigger mode corresponds in function to pressing the
from the front panel, except that it is initiated by sending a bus trigger
command, after selecting BUS as the trigger source.
• The following comand selects the bus trigger source:
34410A/11A User's Guide
hold.
as needed to select HOLD, and press
Press
trigger source:
TRIGger:SOURce IMMediate
The CONFigure and MEASure? commands automatically set the trigger
source to IMMediate.
Refer to the Agilent 34410A/34411A Programmer's Reference for
complete description and syntax for these commands.
TRIGger:SOURce BUS
Then, any of the following commands will initiate readings:
MEASure?
READ?
INITiate
Refer to the Agilent 34410A/34411A Programmer's Reference for
complete description and syntax for these commands.
Features and Functions
to display TRIGGER.
.
key
2
71