Triggering the Multimeter

The multimeter's triggering system allows you to generate triggers either
manually or automatically, take multiple readings per trigger, and insert a
delay before each reading.
From the front panel, you can trigger the multimeter using a single
manual trigger or auto–triggering. Single triggering takes one reading each
time you press
triggering takes continuous readings at the fastest rate possible for the set
configuration (e.g. selected measurement function, auto–zero setting,
resolution setting, etc.).
From the remote interface, triggering the multimeter is a three–step
process: configure measurements, identify trigger source and initiate the
trigger.
From the rear panel, an Ext Trig terminal accepts hardware triggers.

Selecting a Trigger Source

You must specify the source from which the multimeter will accept a
trigger. The power–on default trigger source is auto–triggering, from the
front panel.
• In local operation, the multimeter will accept single triggers by pressing
• Remote Interface Operation: The following command selects the trigger
34410A/11A User's Guide
, or upon receiving an external trigger signal. Auto
1 First, you must configure the multimeter for the measurement by
selecting the function, range, resolution, etc.
2 Second, you must specify the source from which the multimeter will
accept a trigger. Choices are – a software (bus) trigger from the
remote interface, a hardware trigger from the Ext Trig terminal, or an
immediate internal trigger (default trigger source).
3 Third, you must ensure that the multimeter is ready to accept a
trigger from the specified source (called the wait–for–trigger state).
the
key (see
"Single Triggering"
using the rear panel Ext Trig terminal (see
page 76), or continuously take readings in auto trigger mode (see
Triggering"
on page 70).
source from the remote interface:
TRIGger:SOURce {BUS>IMMediate>EXTernal}
Refer to the Agilent 34410A/34411A Programmer's Reference for
complete description and syntax for this command.