Triggering the Multimeter

From the front panel (Local mode), the multimeter always auto–triggers.
Auto triggering takes continuous readings at the fastest rate possible for
the selected measurement configuration.
From the remote interface, triggering the multimeter is a three–step
process:
Immediate Triggering
The immediate triggering mode is available from the remote interface only.
In the immediate trigger mode, the trigger signal is always present. When
you place the multimeter in the wait–for–trigger state, the trigger is issued
immediately. This is the default trigger source for remote interface
operation.
• Remote Interface Operation: The following command selects the immediate
Software (Bus) Triggering
The bus trigger mode is available from the remote interface only.
The bus trigger mode is initiated by sending a bus trigger command, after
selecting BUS as the trigger source.
• The TRIGger:SOURce BUS command selects the bus trigger source.
34405A User's and Service Guide
1 Configure the multimeter for the measurement by selecting the
function, range, resolution, and so on.
2 Specify the multimeter's trigger source. Choices are a software (bus)
trigger from the remote interface or an immediate internal trigger
(default trigger source).
3 Ensure that the multimeter is ready to accept a trigger from the
specified source (called the wait–for–trigger state).
trigger source:
TRIGger:SOURce IMMediate
The CONFigure and MEASure? commands automatically set the trigger
source to IMMediate.
Refer to the Agilent 34405A Programmer's Reference for complete
description and syntax for these commands.