Error Messages - Konica Minolta CM-2600d Instruction Manual

Spectrophometer
Hide thumbs Also See for CM-2600d:
Table of Contents

Advertisement

Error Messages

The following messages may appear while you are using the instrument. If such messages appear, take
the necessary actions shown in the table below. If the trouble does not go away in spite of taking action,
contact a KONICA MINOLTA SENSING-authorized service facility.
M
emo
Messages that may be displayed on the LCD are given in the table below. For communication error check
codes, refer to the separate document.
MessageCorrective Action
OK02
LOW ILLUMINATION
OK10
MAKE
WHITE
TION !
ER01
MEMORY FULL
ER02
LOW BATTERY
ER05
ERROR IN FLASHING
ER07
ZERO CALIBRATION INCOM-
PLETED
ER11
FAILED IN CALIBRATION
E-108
Symptom/Possible Cause
Amount of light from the xenon
lamp has dropped to 65% of its
initial level.
• Deterioration of xenon lamp
• Dirt on integrating sphere
The reliability of the white cali-
CALIBRA-
bration data has dropped because
a certain period of time has
passed since the last white cali-
bration.
When operating in COND mode
or TASK mode, the instrument
has a total of 1700 measurement
and color difference target data
stored. (700 in the "defined in
COND." mode.).
(When the instrument is powered
by batteries)
• Battery voltage drop
Xenon lamp does not flash prop-
erly.
• Xenon lamp has blown.
• Faulty flash circuit
• Faulty sensor
Zero calibration has not been
performed.
Zero calibration or white calibra-
tion has not been performed cor-
rectly.
The Target Mask is not correct.
Corrective Action
Clean the integrating sphere as
explained in "Cleaning Each
Part" (page E-17). If the problem
still remains, contact a KONICA
MINOLTA
SENSING-author-
ized service facility.
Performing white calibration is
recommended.
Delete some data for securing nec-
essary
memory
space.
memory of this instrument can
store up to 1700 (700 in the
"defined in COND." mode)
pieces of data in total. This
amount includes the sample data
for the COND and TASK modes
as well as color difference target
data.
Replace the batteries with new
ones.
Contact a KONICA MINOLTA
SENSING-authorized service fa-
cility.
Perform zero calibration.
For zero calibration, the speci-
men measuring port must be di-
rected into the air. For white
calibration, the White Calibra-
tion Plate must be used.
Attach the correct Target Mask.
The