Error Messages - Konica Minolta CHROMA METER Instruction Manual

Table of Contents

Advertisement

Error Messages

When the instrument does not operate properly as expected, it displays an error message
on the LCD screen. The following table shows possible error messages, their meaning
(cause), and countermeasures.
Error message
Replace the
1
battery.
2
Input value outside
range.
3
Specify the date
and time.
4
Memory full.
5
Measurement data
not found.
6
Target data does
not exist.
7
Backup battery has
run out.
Meaning (cause)
The battery voltage is low.
The value you entered is
out of the available range.
The instrument is started for
the first time
and the date and time have
not been set.
You attempted to save more
than 1,000 measurement
results.
You attempted to check the
measurement result or
perform target setting or
user calibration when there
is no measurement data/
stored data in the
instrument.
You attempted to set a
target when there is no
measurement data/stored
data in the instrument.
The backup battery voltage
dropped, resulting in the
corruption of setting data.
103
Countermeasure
• Set the power switch to OFF
( side) and replace the
batteries with new ones or
use the AC adapter.
• Enter a correct value.
Be sure to enter a value that
satisfies all the conditions of
the value input rule.
* Value input rule (Refer to
page 70.)
• Set correct date and time.
• Follow the guidance and
overwrite the data or delete
unnecessary measurement
results.
• Perform measurement or
select measurement data and
then try the setting again.
• Perform measurement or
select measurement data and
then try the setting again.
• Charge the backup battery.
The battery is fully charged in
about 20 hours.
When the battery is fully
charged, configure the setting
of the instrument again.
• If the problem persists,
contact a KONICA MINOLTA-
authorized service facility.