Water-Saver System - Getinge 600 Series Installation Manual

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Optional Equipment

WATER-SAVER SYSTEM

5–22
The MP-129F Water-Saver is an add-on system that incorporates a water
ejector with a pump and a water reservoir to recirculate the water that
creates the vacuum necessary for part of the sterilization cycle.
Installation of the water-saver consists of:
• Locating, mounting, and anchoring the unit.
• Connecting the cold water supply tubing.
• Connecting the overflow drain line
• Connecting the water-saver to the sterilizer piping (see Figure 5–19 on
page 5–23).
• Connecting the electrical power to the water-saver I/F Box (see
Figure 5–20 on page 5–24).
• Connecting the water-saver to the sterilizer controls (see Figure 5–20
on page 5–24).
The customer must provide a proper drainage system, cold water supply,
and electrical service. Refer to MP–129 (E & F) WATER-SAVER Installation
Instructions (Pub. No. 61301605429).
WARNING
Turn off all electricity, water, and steam to the sterilizer before starting the
installation of the water-saver. Touching live electrical connections and/or
exposure to live steam could cause serious injury.
INS 61301607027 Rev A US