UTILITY REQUIREMENTS

(units with steam boiler)
Steam Supply
(units without steam boiler)
Cold Water Supply
Cold Water
Shutoff
115 or 230 Vac
Single-Phase
Booster Pump
(option)
INS 61301607027 Rev A US
The sterilizer requires steam, cold water and electrical supplies as specified
on the roughing-in drawing. Sterilizers with an integral boiler require a hot
water supply (instead of steam) and three-phase electrical supply as well.
FIGURE 4–19. UTILITY CONNECTIONS
Hot Water Supply
Steam
Hot Water
Shutoff
115 Vac
Shutoff
Power Box
UPS
(option)
Integral
Transformer
(option)
Heater Voltage
Steam Boiler
(option)
600 Series Steam Sterilizers
Single-Phase
Power Source
Universal
Voltages
Transformer
> 380 Vac
(option)
UPS
230 Vac
(option)
Power Source
Mains
Disconnect
Circuit
Breaker
Three-Phase
Mains
Disconnect
Circuit
Breaker
A07027-BW
4–19