Adjust mounting bolts
only if both sides
are not equidistant
INS 61301607027 Rev A US
FIGURE 4–15. CHECKING FOR SQUARENESS
Measure both sides
with try square
Dimension A =
5.625 in. (143 mm)
6. Refer to Figure 4–15. Using a try square (or tape measure), measure the
distance from the front edge of the mounting plate to the bend in the
mounting bracket. Confirm that this dimension is 5.625 in. (143 mm).
Measure both the left and right sides of the mounting plate.
7. If the measurements on the left and right side of the mounting plate are
equidistant, the mounting plate is square and no adjustment is
required. If not, loosen the bolts that secure the mounting bracket to
the plate and adjust the mounting plate as required.
8. If not already installed, insert the alignment pins into the threaded holes
in the bottom of the door plate.
9. Lower the door.
A
600 Series Steam Sterilizers
A
A07027-CC
4–15