c. If the mains voltage is 380 Vac or above, connect the UPS between
Universal Transformer 8A and Power Box 1A (see Figure 5–18).
FIGURE 5–18. UPS CONNECTION (OTHER VOLTAGES)
380/400/415/440/460/480 Vac
Single-Phase Supply
L1 L2
8A
1TB
1
2
3
4
UNIVERSAL
TRANSFORMER
1TB
1
2
3
L
N
115 Vac
1A
POWER BOX
10A
UNINTERRUPTIBLE
POWER SUPPLY
(115 Vac Model)
A08205-D
3. Connect the UPS to the mains disconnect.