600 Series Steam Sterilizers
c. If the mains voltage is 380 Vac or above, connect the UPS between
Universal Transformer 8A and Power Box 1A (see Figure 5–18).
FIGURE 5–18. UPS CONNECTION (OTHER VOLTAGES)
380/400/415/440/460/480 Vac
Single-Phase Supply
L1 L2
8A
1TB
1
2
3
4
UNIVERSAL
TRANSFORMER
1TB
1
2
3
L
N
115 Vac
1A
POWER BOX
10A
UNINTERRUPTIBLE
POWER SUPPLY
(115 Vac Model)
A08205-D
3. Connect the UPS to the mains disconnect.
INS 61301607027 Rev A US