Steam Supply Quality - Getinge 600 Series Installation Manual

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

STEAM SUPPLY QUALITY

INS 61301607027 Rev A US
It shall be the customer's responsibility to provide condensate-free steam
between 97% and 100% saturated vapor. Failure to comply with this
requirement may result in decreased sterilizer function, performance, and
reliability.
Two methods which may assist in meeting this requirement are:
• Installing a thermodynamic steam trap and associated piping as shown
in Figure 7–1.
• Installing an MP-106 Steam Conditioning Package (Figure 7–2 on page
7–4) which may be purchased from Getinge.
FIGURE 7–1. TYPICAL STEAM SUPPLY PIPING
Pressure Regulator
Separator
To Drain
Manual
Shutoff Valve
Building Steam Supply
To Vent
Valve
Steam Trap
600 Series Steam Sterilizers
Pressure
Relief Valve
97 - 100%
Saturated Steam
to Sterilizer
Pressure
Gauge
WS-0227A
7–3