LEVELING THE STERILIZER
CHAMBER
INS 61301607027 Rev A US
Once the sterilizer is positioned, check the alignment and leveling of the
chamber. The unit must be level to run properly and align with the loading
carts.
NOTE
If power is available, connect the power to the input terminals inside the
sterilizer Power Box and use the Open Door switch to open the door.
Otherwise, the door must be opened by manually turning the lift screw. Use
a spanner wrench to turn the pulley wheel at the base of the lift screw.
To level the sterilizer chamber:
1. If power is available:
a. Connect 115 Vac to 1TB in Power Box 1A (see "Connecting the
Electrical Power" on page 4–22
b. Turn ON the Power Box circuit breaker.
FIGURE 4–10. STERILIZER CONTROL PANEL
c. Press the CONTROLS ON/OFF switch on the Control Panel.
d. Turn the CONTROLS DISABLE/ENABLE switch to the ENABLE
position.
e. Press the Open Door switch.
600 Series Steam Sterilizers
)
.
CONTROLS
OFF
ON
CONTROLS
ON/OFF Switch
CONTROLS
DISABLE
ENABLE
CONTROLS
ENABLE/DISABLE
Switch
Close Door
Switch
Open Door
Switch
A07027-BU
4–9