Getinge 600 Series Installation Manual page 63

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

See table below for current ratings
and wire sizes.
K
Electric – 4-wire with ground
Service Entrance for
Three-Phase Power Supply
VOLTAGE
208V
240V
480V
600V
INS 61301607027 Rev A US
NOTE
The 208 Vac version of the CAS-45A has a fuse box mounted on the right
side of the electrical cabinet.The customer three-phase connection is
routed through six 100A fuses before it is connected to the heating element
contactors. For all other voltages, the customer three-phase connection is
routed to the heating element contactors.
FIGURE 5–3. CAS-45A STEAM BOILER ELECTRICAL CONNECTIONS
CURRENT
POWER
125.1 A
45 KW
108.4 A
45 KW
54.2 A
45 KW
43.4 A
45 KW
Be sure wires are clean and bright to
ensure good electrical contact.
Three-Phase Customer
Supply Connection
(torque to 110 in-lbs.)
K
WIRE SIZE
CONDUIT
(90°C Copper)
2 AWG
3 AWG
8 AWG
8 AWG
600 Series Steam Sterilizers
Ground
Connection
Wiring to Fuse Box
(208 Vac model only)
SIZE
1¼"
1¼"
¾"
¾"
A07027-R
5–5