Getinge 600 Series Installation Manual page 50

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Installation Instructions
Integral
Transformer
Box 9A
230 Vac ±10%
Single-Phase Supply
9A
INTEGRAL
TRANSFORMER
BOX
4–24
230 VAC INSTALLATION
Integral Transformer 9A is a factory-installed option used to step down the
230 Vac customer supply to the 115 Vac required by the sterilizer.
Connect the 230 Vac 50/60 Hz single-phase supply circuit to terminals L
(hot), N (neutral), and protective ground on terminal block 1TB inside the
Integral Transformer Box (see Figure 4–23).
FIGURE 4–23. CONNECTING THE POWER SOURCE (230 VAC)
L
N
1
2
3
1TB
1
2
3
2TB
Output to
Power Box 1A
(115 Vac)
2TB
1
2
3
Terminal Block
1
2
3
L
N
1TB
Customer Supply
(230 Vac)
1A
POWER BOX
115 Vac
INS 61301607027 Rev A US
1TB
1TB
1
2
3
L
N
A08205-H