Getinge 600 Series Installation Manual page 37

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

INS 61301607027 Rev A US
2. Refer to Figure 4–12 on page 4–12. Place the level vertically against the
chamber wall and adjust the feet to level the chamber from side-to-
side. The minimum distance from the finished floor must be a minimum
of 2.25 inches (57 mm) as shown in Figure 4–5 on page 4–5.
3. Place the level vertically against the door gasket headring and adjust
the feet to level the chamber from front-to-back. The minimum
distance from the finished floor must be a minimum of 2.25 inches (57
mm) as shown in Figure 4–5 on page 4–5.
NOTE
If a loading car and transfer carriage will be used with the sterilizer, be sure
to maintain the proper height adjustment shown in Figure 4–6 on page 4–6.
4. Once the chamber is level, tighten the locknuts on the leveling feet.
5. Raise the sterilizer door by pressing the Close Door switch or turning
the lift screw pulley clockwise.
600 Series Steam Sterilizers
4–11