Isolating Device - Getinge 600 Series Installation Manual

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Safety

ISOLATING DEVICE

Power Box
2–4
The sterilizer is fitted with a main switch on the power box (located on the
right side of the sterilizer). This switch removes primary power (120 Vac)
from the sterilizer.
Sterilizers with a steam boiler have a three-phase heating circuit that
requires a separate mains disconnect. The power switch on the boiler turns
the boiler control circuits ON and OFF.
FIGURE 2–3. CIRCUIT BREAKER LOCATION
Main Switch
Steam Boiler
(option)
NOTE
The CONTROLS ON/OFF switch turns the sterilizer controls ON and OFF
but does not remove primary power.
The CONTROLS DISABLE/ENABLE switch disables the door motor circuits
to prevent door movement during routine maintenance.
The control voltage for the optional steam boiler is provided by the sterilizer
power box.
Heater Voltage
(208/240/480/600 Vac
Three-Phase)
INS 61301607027 Rev A US
Mains
Disconnect
Breaker
A07027-K