Connecting The Customer Supply - Getinge 600 Series Installation Manual

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

CONNECTING THE CUSTOMER
SUPPLY
INS 61301607027 Rev A US
The primary power connection procedure is dependent on the supply
voltage:
• 115 Vac—power is connected to the Power Box.
See "115 Vac Installation" below.
• 230 Vac—power is connected to the factory-installed Integral
Transformer which steps down the voltage to the 115 Vac required by
the sterilizer. See "230 Vac Installation" on page 4–24.
• 380 Vac and above—power is connected to the Universal Transformer
which steps down the voltage to 115 Vac. The Universal Transformer is
installed and wired by the customer. See "Universal Transformer" on
page 5–16.
NOTE
If the sterilizer has an uninterruptible power supply (UPS), install the UPS
before connecting power to the sterilizer. see "Uninterruptible Power
Supply (UPS)" on page 5–18 for electrical connections.
115 VAC INSTALLATION
Connect the 115 Vac 50/60 Hz single-phase supply circuit to terminals L
(hot), N (neutral), and protective ground on terminal block 1TB inside the
Power Box (see Figure 4–22).
FIGURE 4–22. CONNECTING THE POWER SOURCE (115 VAC)
J5
J4
J3
J2
J1
600 Series Steam Sterilizers
Customer Supply
(115 Vac ±10% Single-Phase)
Power Box 1A
(with Cover Open)
Terminal Block
1TB
A08205-K
4–23