Technical Details; Cleaning And Care - Sanitas SBM 03 Operating Instructions Manual

Table of Contents

Advertisement

Available languages

Available languages

• inflation takes longer than 3 minutes (Er6 appears in
the display),
• the cuff is too tight or too loose (Er3 or Er4 appears in
the display),
• the blood-pressure values are unusually high or low
(Er1 or Er2 appears in the display),
• the inflation pressure is higher than 300 mmHg (Er5
appears in the display),
• there is a system or device error (Er0, Er7, Er8, Er9,
ErA or ErB appears in the display)
• the batteries are almost flat; the batteries must be
replaced.
In the above cases, you must repeat the measure-
ment.
Reinsert the batteries if necessary, or else replace
them.

8. Cleaning and Care

• Clean your blood pressure computer carefully using
only a slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise
liquid can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement
method:
Measurement range
Precision of
pressure reading
Measurement
uncertainty
Memory
Dimensions
Weight
Cuff size
Admissible operating +5 °C to +40 °C, 40-80%
temperature
Admissible
storage temperature
Power supply
Battery life
Accessories
17
SBM 03
Oscillating, non-invasive blood
pressure measurement on the
wrist
Pressure 0-300 mmHg, heart
rate 40-180 beats/minute
Systolic ± 3 mmHg /
diastolic ± 3 mmHg
pulse ± 5% of displayed value
Max. admissible standard
deviation according to clinical
examination: systolic 8 mmHg /
diastolic 8 mmHg
1 x 60 storable data lines
L 84 x W 62 x H 30 mm
Approx. 140 g
For a wrist circumference of
approx. 140 to 195 mm
relative humidity
-20 °C to +55 °C, 10-95%
relative humidity
2 x 1.5 V Micro
(Alkaline Type LR 03)
For 250 measurements
Storage box, instructions for
use, 2 x AAA batteries