" Lo ". In such cases, you should seek medical assistance and check
the accuracy of your procedure.
The limit values for the technical alarm are factory set and cannot
be adjusted or deactivated. These alarm limit values are accorded
second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be reset.
The signal shown on the display will disappear automatically after
about 8 seconds.

8. Cleaning and Care

• Clean your blood pressure monitor carefully using a slightly damp
cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid can
penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
SBC 23
Measurement
Oscillometric, non-invasive blood pressure
method
measurement on the wrist
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
60 memory spaces
Dimensions
L 85 mm x W 64 mm x H 24 mm
Weight
Approx. 104 g (without batteries)
Cuff size
140 to 195 mm
Permissible operating
+10 °C to +40 °C, ≤ 85 % relative air humidity
conditions
(non-condensing)