Error
Possible cause
mes-
sage
The inflation
pressure is
higher than
300 mmHg.
There is a sys-
tem error.
The batteries are
almost empty.

11. Disposal

Repairing and disposing of the device
• Do not repair or adjust the device yourself. Proper operation
can no longer be guaranteed in this case.
• Do not open the device. Failure to comply will invalidate the
warranty.
• Repairs must only be carried out by Customer Services or
authorised retailers. Before making a claim, please check the
batteries first and replace them if necessary.
Solution
Please take another measure-
ment to check whether the cuff
can be correctly inflated.
Make sure that neither your
arm nor other heavy objects
are pressing on the line, and
that the line is not bent.
If this error message appears,
please contact Customer
Services.
Insert new batteries into the
device.
• For environmental reasons, do not dispose of the
device in household waste at the end of its service life.
Dispose of the device at a suitable local collection or
recycling point in your country. Dispose of the device in
accordance with EC Directive – WEEE (Waste Electrical and
Electronic Equipment). If you have any questions, please
contact the local authorities responsible for waste disposal.
Disposing of the batteries
• The empty, completely flat batteries must be disposed of
through specially designated collection boxes, recycling
points or electronics retailers. You are legally required to
dispose of the batteries.
• The codes below are printed on batteries
containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.

12. Technical specifications

Model No.
SBM 22
Measurement
Oscillometric, non-invasive blood pressu-
method
re measurement on the upper arm
Measurement
Cuff pressure 0-300 mmHg,
range
Systolic 50-280 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-199 beats/minute
Display accuracy
Systolic ± 3 mmHg, Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown