6. Saving, retrieving and deleting results

• The results of each successful measurement are stored to-
gether with date and time. With more than 60 items of meas-
ured data, the earliest items of data measured are lost.
• Select the desired user memory by pressing the MEM but-
ton and then the "+" button. Press the MEM button again to
display the average of all saved measurements in the user
memory. Press the MEM button again to display the average
day measurement for the last 7 days. (day: 5.00 a.m. – 9.00
a.m., display A). Press the MEM button again to display the
average night measurement for the last 7 days
(night: 5.00 p.m. – 9.00 p.m., display P). If you continue press-
ing the MEM button, the latest individual results are displayed
with date and time.
• To clear the memory, press the MEM button, then the display
shows no. 1. Press the "+" button to select the user memory
and confirm by pressing MEM again. Press and hold the "+"
and
keys simultaneously for 5 seconds (the display will
show CLA).
• If you wish to change the user memory, see "Select memory".
7. Error messages / troubleshooting
In case of faults, the E_ message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg E2.
2. blood pressure is unusually high or low E3.
3. you move or talk during the measurement (in addition to E_,
the cardiac arrhythmia icon
4. the cuff tube is not inserted properly E1.
5. the cuff is inflated for longer than 25 seconds E1.
appears in the display E3),
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.

8. Cleaning and Care

• Clean your blood pressure computer carefully using only a
slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement
method
Measurement range
Precision of
pressure reading
Measurement
uncertainty
Memory
Dimensions
Weight
Cuff size
15
SBC 27
Oscillating, non-invasive blood
pressure measurement on the wrist
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
pulse 40 –199 beats / minute
Systolic ± 3 mmHg /
diastolic ± 3 mmHg /
pulse ± 5 % of displayed value
Max. admissible standard deviation
according to clinical examination.
systolic 8 mmHg / diastolic 8 mmHg
2 x 60 storable data lines
L 72 x W 69 x H 29,2 mm
Approx. 115 g
For a wrist circumference of approx.
135 to 195 mm