20 – Thermographic measurement techniques
20.2.1

Finding the emissivity of a sample

20.2.1.1

Step 1: Determining reflected apparent temperature

Use one of the following two methods to determine reflected apparent temperature:
20.2.1.1.1
Method 1: Direct method
1
Look for possible reflection sources, considering that the incident angle = reflection angle (a
= b).
10588903;a1
Figure 20.1 1 = Reflection source
2
If the reflection source is a spot source, modify the source by obstructing it using a piece if
cardboard.
10589103;a2
Figure 20.2 1 = Reflection source
128
Publ. No. T559382 Rev. a358 – ENGLISH (EN) – June 23, 2009