DC CURRENT
Ranges: 430µA,43mA,430mA,10A
Resolution: 100nA
Accuracy: ±(0.5%rdg + 1dgt) on µA,mA ranges
±(2.0%rdg + 1dgt) on 10A range
Burden voltage: 1.4V on all ranges, execpt
Input protection:0.5A / 250V fast blow ceramic fuse
AC CURRENT (50Hz - 1KHz)
Ranges: 430µA,43mA,430mA,10A
Resolution: 100nA
Accuracy: ±(1.0%rdg + 2dgts) on µA,mA ranges
±(2.5%rdg + 2dgts) on 10A range
Burden voltage: 1.4V on all ranges, execpt
Input protection:0.5A / 250V fast blow ceramic fuse
RESISTANCE
Ranges: 430W ,4.3KW ,43KW ,430KW ,4300KW ,43MW
Resolution: 100mW
Accuracy: ±(0.3%rdg + 3dgts) on 430W range
±(0.3%rdg + 2dgts) on 4.3KW to 4300KW ranges
±(1.5%rdg + 4dgts) on 43MW range
Audible indication: <50W ±30W
Open circuit volts: 0.4Vdc
Overload protection: 500VDC or AC rms
1.5V on 10A range
10A / 600V fast blow ceramic fuse
1.5V on 10A range
10A / 600V fast blow ceramic fuse