SpeciFicaTiONS

cpap
Size ........................................................................................................... 4.2"H x 6.5"W x 6.9"D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC .................................................................................................................. 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Bilevel Pressure Range ...........................................................................................................................................3-25 cmH
Operating Temperature Range ....................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range ..................................................................... -40°F to +158°F (-40°C to 70°C)
Storage & Transportation Humidity Range ...............................................................................0% to 95% RH non-condensing
Maximum Limited Pressure ...........................................................................................................30 cmH
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10 mgH
Size: ......................................................................................................2.6" H x 6.3" W x 8.4" D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305°F (152°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Temperature Range ....................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter Specifications
Standard Filter .........................................................................................................................................> 3.0 micron particles
Optional Fine Particle Filter .....................................................................................................................> 0.3 micron particles
Dynamic Short-Term pressure accuracy with and without DV5HH Humidifier (per
iSO 17510-1:2007)
Waveform
Volume [mL]
Sinusoidal
cycle with I:E =
1:1
A-DV56
1. Trigger Sensitivity is not set correctly.
2. Backup rate (time) is too high, the patient is
breathing slower than the backup rate.
Breath Rate [min-1]
500
10, 15, 20
remedy
1. Adjust the inspiratory and expiratory trigger
sensitivity to adjust triggering for the patient.
2. Reduce Backup Rate setting.
Set Pressure
[cmH
O]
2
3.0
9.0
14.0
20.0
25.0
2
O under normal use
2
O/l air
2
Pressure Accuracy
[Pk-Pk cmH
O]
2
0.5
0.5
1.0
1.0
1.0