Table of Contents Introducing the ALERT J4X.ec Introduction Operating Conditions ALERT J4X.ec Breath Test Operations Portable Breath Analysis System Rocker Switch Function Initialization Sequence Taking a Passive Test Taking a Direct Test Printing the Test Results Last Test Recall Operations Manual...
Introducing the ALERT J4X.ec Initialization Sequence Introduction Each time the ALERT J4X.ec unit is turned on it will go through a series of self diagnostic checks. First “8.8.8.8” The ALERT J4X.ec is a portable breath alcohol tester with will appear on the display, followed by the current date printer interface and PC communications capability.
Last Test Recall to stop. The ALERT J4X.ec unit permits the recall of the data for The flow of breath will be automatically monitored and a the last breath test conducted for viewing on the display sample of deep lung air will be captured for analysis.
Warnings and Special Precautions Err 7 Test memory log full. Connect ALERT J4X.ec to PC and run program to read and clear log. Breath coming from the lungs must not be contaminated with smoke particles or "mouth alcohol" from a recent Err 8 Calibration range low.
Service Battery Pack The battery pack for the ALERT J4X.ec unit consists of a rechargeable nickel metal hydride (NiMH) battery pack. Battery Level Indicator Each time the ALERT J4X.ec unit is turned on the unit will go through a series of self diagnostic checks which includes the readout of battery voltage on a bar graph.
Specifications Accessory and Disposable Products Size 15 cm x 8.8 cm x 3.5 cm 94-000140 TOXITEST Weight 320 grams Breath Alcohol Simulator Sensor Electro-chemical (fuel cell) 95-000132 MOUTHPIECE (Round) Specificity No response to ketones Individually Wrapped (25/Bag) Ambient Range -5 ºC to 50 ºC 95-000310 STANDARD SOLUTION (500 mL/.100 %) Purge Cycle (Initial Test)
Need help?
Do you have a question about the ALERT J4X.ec and is the answer not in the manual?
Questions and answers