Warnings and Cautions
Avoid explosive or flammable
gases
It has been determined that the SAM
500P is safe to use with oxygen mask
delivery systems. However, to avoid
the risk of an explosion, it is strongly
advised that you do NOT use the SAM
500P in the vicinity of explosive gases,
including flammable anaesthetics or
concentrated oxygen.
6
Caution
Correct placement of the electrode
pads
Proper placement of the SAM 500P's
electrode pads is critical. You must strictly
observe the instructions shown in the
Emergency User Guide and on the device.
Wrong placement, or the presence of
air, hair, surgical dressings or medicine
patches between the pads and the skin,
could reduce defibrillation effectiveness.
Slightly red skin after shock therapy
is normal.
Do not touch the patient during
analysis
Touching the patient during the analysis
phase of treatment can cause interference
with the diagnostic process. Avoid contact
with the patient while analysis is being
carried out. The device will instruct you
when it is safe to touch the patient.