Guidance and manufacturer's declaration – electromagnetic immunity
The SAM 500P is intended for use in the electromagnetic environment specified below. The customer or user
of the SAM 500P must assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
IEC/EN 61000-4-2
± 8 kV air
Electrical fast
±2kV for power supply lines
transient/burst
IEC/EN 61000-4-4
±1 kV for input/output lines
Surge
±1kV differential mode
IEC/EN 61000-4-5
±2 kV common mode
Voltage dips, short
<5 % Ut (>95 % dip in Ut)
interruptions and
for 0.5 cycle
voltage
40 % Ut (60% dip in Ut)
variations on
for 5 cycles
power supply
70 % Ut (30 % dip in Ut)
input lines
for 25 cycles
IEC/EN 61000-4-11
<5 % Ut (>95 % dip in Ut)
for 5 sec
Power-frequency
(50/60 Hz)
3 A/m
magnetic field
IEC/EN 61000-4-8
Note: Ut is the a.c.mains voltage prior to application of the test level
Compliance
Electromagnetic environment
level
– guidance
Complies
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
Complies
humidity should be at least 30%.
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Not Applicable
Power frequency magnetic fields
should be at levels characteristic
3A/m
of a typical location in a typical
commercial or hospital environment.
37