2. Unpack the Pad-Pak. Retain the
packaging in case you need to return
the Pad-Pak to HeartSine. Place the
SAM 500P on a flat surface. Insert
the Pad-Pak into the SAM 500P (see
the illustration below). Listen for the
"click" sound and ensure both tabs
are fully engaged.
3. If required, the SAM 500P will run a
self test routine. The action arrows
will flash during this process. On
successful completion of the self-test
routine, the green status indicator
(see 'SAM 500P Overview' on page
13) will blink. If so, your SAM
500P is ready for use.
4. Turn on the SAM 500P by pressing
on the front panel to check that
the device is operating correctly.
Listen for the voice prompts but do
follow them. Make sure that no
NOT
warning messages are played.
Notice: Do NOT pull the green
tab on the Pad-Pak. If you have
opened the electrode drawer,
you may have to replace
your Pad-Pak.
Only turn the SAM 500P on
ONCE. If you turn it on and off
repeatedly, you will exhaust
the batteries prematurely
and you may need to replace
the Pad-Pak.
5. Turn off the SAM 500P by pressing
on the front panel. Check the
status indicator (see 'SAM 500P
Overview' on page 13) is flashing
green. If you have heard no warning
messages and the status indicator
is flashing green, the device is
ready for use.