The SAM 500P ECG arrhythmia analysis algorithm performance is summarised in the
table below:
Rhythm Class
Shockable Rhythm:
Ventricular Fibrillation (VF)
Shockable Rhythm:
Ventricular Tachycardia (VT)
Non-Shockable Rhythm:
Combined Non-Shockable Rhythms
Non-Shockable Rhythm:
Asystole
*No error to measure
ECG Test
Required
Sample
Performance
Size
Specifications
(seconds)
13341
Sensitivity > 90%
1946
Sensitivity > 75%
286056
Specificity > 95%
10839
Specificity > 95%
90% One-
Performance
Sided Lower
Results (%)
Confidence
Limit
96.97
96.72
91.36
90.25
99.04
99.01
100*
99.97*